Anaesthetic Gas Scavenging Systems

Two Types of AGSS Terminal Units: ISO 9170-2 and Power-Device Location

ISO 9170-2 distinguishes two AGSS terminal types according to whether the power device is upstream or downstream. Understand how to identify the required terminal and probe in a hospital equipment schedule.

AI-generated editorial equipment image: Two Types of AGSS Terminal Units: ISO 9170-2 and Power-Device Location
Two Types of AGSS Terminal UnitsBWC Medgas · AI-generated equipment illustrationAI-generated editorial equipment image. Illustrative context, not a product specification or project record.

Hospital operating departments may require terminal connections for anaesthetic gas scavenging disposal systems (AGSS). These can be mistaken for ordinary suction or medical vacuum connections because both systems are associated with moving gases away from clinical equipment.

The ISO 9170-2:2008 standard provides an important distinction: it describes two AGSS terminal-unit types, defined by whether the power device is located upstream or downstream of the terminal.

That distinction belongs in the hospital's equipment schedule before an AGSS terminal is purchased.

What is the scope of ISO 9170-2?

The ISO 9170-2:2008 record covers terminal units intended for use in anaesthetic gas scavenging disposal systems under ISO 7396-2.

The standard specifies two terminal types according to the location of the power device relative to the terminal unit. It also describes requirements and dimensions for the mating counterpart (probe) of the type-specific connection.

Therefore an order for “AGSS wall outlet” is incomplete if the project does not identify which terminal type and matching connector are required.

Why does the location of the power device matter?

The power device is part of the scavenging arrangement. Its position relative to the terminal affects which terminal configuration is specified by the engineered system.

The published abstract establishes the existence of two types but does not provide a simple universal replacement rule between them. The hospital's approved system drawings must determine the type.

A buyer should never infer the required type from the position of the terminal faceplate or from a photograph taken in another facility.

Does ISO 7396-2 cover the same thing?

ISO 7396-2:2007 concerns anaesthetic gas scavenging disposal systems, including design and performance of the broader disposal arrangement. It is not merely a product standard for an individual terminal.

By contrast, ISO 9170-2 addresses the particular terminal units in AGSS. These two scopes support each other but are not interchangeable specifications.

The purchaser should identify both the intended system and the actual terminal/probe package.

Is an AGSS terminal the same as a medical vacuum outlet?

No. ISO 9170-1:2017 addresses medical gas and vacuum terminal units and explicitly points to ISO 9170-2 for AGSS terminal requirements.

A medical vacuum outlet is not automatically a substitute for an AGSS connection. The gas-service identification, type-specific mating dimensions and system design must be preserved.

This is not an academic distinction: a purchase mistake could cause a mismatched connector or delay the installation of a critical hospital room.

What belongs in an AGSS outlet quotation?

For accurate equipment sourcing, the buyer should provide:

  • The project designation AGSS and the adopted engineering specification.
  • The scavenging disposal system drawing identifying the intended terminal type.
  • Whether the power-device arrangement is upstream or downstream, as specified by the system designer.
  • The type-specific terminal and mating probe details.
  • The mounting format and relevant connection dimensions.
  • The number of terminals and any explicitly included accessories.

A supplier should not substitute a standard medical vacuum regulator or vacuum terminal without exact project documentation and engineering approval.

Where BWC Medgas fits

BWC provides medical gas and vacuum terminal products and other MGPS components. Those generic categories are not a declaration that a particular BWC item is a compliant AGSS terminal.

For an AGSS component enquiry, supply the exact required terminal type and project drawings so the actual supply scope can be assessed correctly. A generic outlet photograph should never be treated as sufficient evidence.

Discuss a defined AGSS component requirement with BWC Medgas.

Frequently asked questions

Does ISO 9170-2 define two AGSS terminal types?

Yes. The published standard identifies two types according to the location of the power device.

Can ordinary medical vacuum equipment be substituted for AGSS terminals?

Not by assumption. The system scopes and terminal standards differ.

Primary technical references